• Programmable insulation (50 to 1,200 V DC) and dielectric strength (AC/DC) testing • Program up to 32 files of test types, test points (50 steps), and measurement settings • Optional scanner for multipoint automatic testing • Uses the PWM method to generate accurate test voltages that do not depend on the supply voltage • Ramp timer function for increasing or decreasing the applied voltage during dielectric strength testing at user-specified times

Thông tin người bán

HIOKI VIỆT NAM

6 Hoa Ma, Hai Ba Trung, Ha Noi

Hotline: 0917047888

Email: hiokivietnam.vic@gmail.com

Thông số kỹ thuật

■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Functions Program up to 32 files of 50 step test settings.
10 sets each of dielectric strength and insulation test settings, hold, buzzer
Monitor functions Output voltage, detected current, measured resistance, Refresh rate: 2 times/s
Power supply 100 to 120 V, 200 to 240 V AC, (50/60 Hz), 1000 VA max.
Dimensions and mass 320 mm (12.60 in) W × 155 mm (6.10 in) H × 480 mm (18.9 in) D, 18 kg (634.9 oz)
Accessories H.V. Test Lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1

[Withstanding test]

Testing voltage AC/DC 0.2 kV to 5.00 kV AC, 500 VA (max. 30 minutes), 0.2 kV to 5.00 kV DC, 50 VA (continuance)
Voltage setting Digital setting (0.01 kV setting resolution)
Waveform/Frequency Sine wave (5% or less distortion, unloaded), 50/60 Hz selectable
Current measurement 0.01 mA to 100.0 mA, Average rectified display (Digital)
Measurement range 10 mA (0.01 mA resolution), 100 mA (0.1 mA resolution)
Voltmeter Digital: accuracy ±1.5 % f.s. (f.s.=5.00 kV) (Average rectified display)
Decision method Window comparison (digital settings)

[Insulation test]

Rated testing voltage 50 to 1,200 V DC (in 1 V steps)
Rated testing current 1 mA, Short-circuit current: 200 mA or less
Measurement range/accuracy 0.10 to 9999 MΩ, 4 ranges, ± 4 % rdg. (representative values for 0.5 MΩ to 1,000 MΩ)
Decision method Window comparison (digital settings)

[Timer]*Test times may differ from set timer times depending on the load.

Setting range 0.3 to 999 s
Ramp/delay Testing voltage ramp-up, or down, Insulation test delay: 0.1 to 99.9 s

 

Options

Input/Output cords

*The 9615 is bundled

REMOTE CONTROL BOX (SINGLE) 9613

For Start/Stop control, 1.5m (4.92 ft) cord length

 

REMOTE CONTROL BOX (DUAL) 9614

For Start/Stop control, 1.5m (4.92 ft) cord length

 

H.V. TEST LEAD 9615

Red, Black each 1, 1.5 m (4.92 ft) length

PC communication

SAFETY TEST DATA MANAGEMENT SOFTWARE 9267

• PC software for use with HIOKI electrical safety testers • Inspect, record and save measurement data according to electrical safety standards controlling electrical safety measuring instruments • Test insulation, withstand voltage (hipot), protective continuity, and leakage current

Learn more

 

RS-232C CABLE 9637

9 pin - 9 pin, cross, 1.8 m (5.91 ft) length

 

GP-IB CONNECTOR CABLE 9151-02

2 m (6.56 ft) length

Other

HIGH VOLTAGE SCANNER 3930

• Supports remote control • For automatic multipoint testing of insulation / withstand voltage • Use with the 3153's program or with general-purpose logic sequencers

Learn more