Setaflash Series 8 ActiveCool Flash Point Tester - 82100-2
Hãng : Stanhope Seta- Anh
Model : 82100-2
Methods : ASTM D1655, ASTM D3278, ASTM D3828, ASTM D7236, ASTM D8174, ASTM E502, IP 523, IP 534, BS EN ISO 3679 & BS 2000-523, CLP Regulations EC No 1272/2008, EPA 1020 B, ISO 3679, ISO 3680 (Obs), SW-846 1020C
The Setaflash Series 8 ActiveCool, closed cup flash point tester provides fast and reliable flash point results with a wide temperature range of -30 to 135 °C with no external cooling required.
With the Series 8, a test takes less than 2 minutes using just a 2 or 4 ml sample size. A simple user interface with colour digital display and touchscreen icons guides you through the straightforward testing process. For easy record keeping the instrument stores 100,000 test results which can be saved and transferred via a USB port.
Designed for precise and accurate tests, barometric pressure is measured by an internal barometer and correction is automatically applied to the test result. Unknown samples can be tested using ramp mode.
Please note, 82100-2 is comparable to ASTM D3278, ASTM D7236, IP 524 and IP 534, however to be fully compliant a gas ignition should be used.
Details
| Temperature range | -30 to 135 °C |
|---|---|
| Temperature selection | °C or °F |
| Ignitor | Electric |
| Heating/cooling method | Peltier cell |
| Temperature ramp mode | Automatic or custom settings (ramp rate up to 10 °C/min) |
| Sample cup material | Aluminium |
| Pressure correction | Automatic |
| Data storage | 1 GB |
| Voltage | 110 to 250 Vac, 50/60 Hz |
| Size (HxWxD) | 30 cm x 34 cm x 38 cm |
| Weight | 8 kg |
| CCCN Code | Tariff 90268020 |